A disposable sampling device to collect volume-measured DBS directly from a fingerprick onto DBS paper

    Gabriel Lenk

    *Author for correspondence:

    E-mail Address: glenk@kth.se

    Department for Micro & Nanosystems, KTH Royal Institute of Technology, Osquldas väg 10, 10044 Stockholm, Sweden

    ,
    Sören Sandkvist

    Department of Clinical Pharmacology, Karolinska University Hospital, 14152 Stockholm, Sweden

    ,
    Anton Pohanka

    Department of Clinical Pharmacology, Karolinska University Hospital, 14152 Stockholm, Sweden

    ,
    Göran Stemme

    Department for Micro & Nanosystems, KTH Royal Institute of Technology, Osquldas väg 10, 10044 Stockholm, Sweden

    ,
    Olof Beck

    Department of Clinical Pharmacology, Karolinska University Hospital, 14152 Stockholm, Sweden

    &
    Niclas Roxhed

    Department for Micro & Nanosystems, KTH Royal Institute of Technology, Osquldas väg 10, 10044 Stockholm, Sweden

    Published Online:https://doi.org/10.4155/bio.15.134

    Background: DBS samples collected from a fingerprick typically vary in volume and homogeneity and hence make an accurate quantitative analysis of DBS samples difficult. Results: We report a prototype which first defines a precise liquid volume and subsequently stores it to a conventional DBS matrix. Liquid volumes of 2.2 µl ± 7.1% (n = 21) for deionized water and 6.1 µl ± 8.8% (n = 15) for whole blood have been successfully metered and stored in DBS paper. Conclusion: The new method of collecting a defined volume of blood by DBS sampling has the potential to reduce assay bias for the quantitative evaluation of DBS samples while maintaining the simplicity of conventional DBS sampling.

    Papers of special note have been highlighted as: • of interest; •• of considerable interest

    References

    • 1 Guthrie R, Susi A. A simple phenylalanine method for detecting phenylkenoturia in large population of newborn infants. Pediatrics 32(3), 338–343 (1963).
    • 2 Lofgren SM, Morrissey AB, Chevalier CC et al. Evaluation of a dried blood spot HIV-1 RNA program for early infant diagnosis and viral load monitoring at rural and remote healthcare facilities. AIDS 23(18), 2549–2466 (2009).
    • 3 Edelbroek PM, van der Heijden J, Stolk LML. Dried blood spot methods in therapeutic drug monitoring: methods, assays, and pitfalls. Ther. Drug Monit. 31(3), 327–336 (2009).
    • 4 Manicke N, Abu-Rabie P, Spooner N, Ouyang Z, Cooks R. Quantitative analysis of therapeutic drugs in dried blood spot samples by paper mass spectrometry: an avenue to therapeutic drug monitoring. J. Am. Soc. Mass Spectr. 22(9), 1501–1507 (2011).
    • 5 Barfield M, Spooner N, Lad R, Parry S, Fowles S. Application of dried blood spots combined with HPLC–MS/MS for the quantification of acetaminophen in toxicokinetic studies. J. Chromatogr. B 870(1), 32–37 (2008).
    • 6 Beaudette P, Bateman KP. Discovery stage pharmacokinetics using dried blood spots. J. Chromatogr. B 809(1), 153–158 (2004).
    • 7 Spooner N, Lad R, Barfied M. Dried blood spots as a sample collection technique for the determination of pharmacokinetics in clinical studies: considerations for the validation of a quantitative bioanalytical method. Anal. Chem. 81(4), 1557–1563 (2009).
    • 8 Amsterdam Pv, Waldrop C. The application of dried blood spot sampling in global clinical trials. Bioanalysis 2(11), 1783–1786 (2010).
    • 9 Deniff P, Spooner N. The effect of hematocrit on assay bias when using DBS samples for the quantitative bioanalysis of drugs. Bioanalysis 2(8), 1385–1395 (2010).
    • 10 Deniff P, Woodford L, Spooner N. Effect of ambient humidity on the rate at which blood spots dry and the size of the spot produced. Bioanalysis 5(15), 1863–1871 (2013).•• Identifying inhomogeneous sample distribution in DBS as problematic when partial punches are taken.
    • 11 ÓMara M, Hudson-Curtis B, Olson K, Yueh Y, Dunn J, Spooner N. The effect of hematocrit and punch location on assay bias during quantitative analysis of dried blood spot samples. Bioanalysis 3(20), 2335–2347 (2011).•• Detailed examination of inhomogeneities in DBS by two different methods.
    • 12 Cobb Z, de Vries R, Spooner N et al. In-depth study of homogeneity in DBS using two different techniques: results from the EBF DBS-microsampling consortium. Bioanalysis 5(17), 2161–2169 (2013).
    • 13 Ren X, Paehler T, Zimer M, Guo Z, Zane P, Emmons GT. Impact of various factors on radioactivity distribution in different DBS papers. Bioanalysis 2(8), 1469–1475 (2010).
    • 14 Denniff P, Spooner N. Volumetric absorptive microsampling: a dried sample collection technique for quantitative bioanalysis. Anal. Chem. 86(16), 8489–8495 (2014).
    • 15 Leuthold LA, Heudi O, Déglon J et al. New microfluidic-based sampling procedure for overcoming the haematocrit problem associated with dried blood spot analysis. Anal. Chem. 87(4), 2068–2071 (2015).
    • 16 Chao TC, Trybala A, Starov V, Das DB. Influence of haematocrit level on the kinetics of blood spreading on thin porous medium during dried blood spot sampling. Colloid Surf. A Physiochem. Eng. Aspects 451, 38–47 (2014).
    • 17 de Vries R, Barfield M, Schmid B et al. The effect of hematocrit on bioanalysis of DBS: results from the EBF DBS-microsampling consortium. Bioanalysis 5(17), 2147–2160 (2013).• Demonstrating a method to estimate the hematocrit of DBS.
    • 18 Capiau S, Stove VV, Lambert WE, Stove CP. Prediction of the hematocrit of dried blood spots via potassium measurements on a routine clinical chemistry analyzer. Anal. Chem. 85, 404–410 (2012).