Design and characterization of a 3D-printed staggered herringbone mixer

    Vedika J Shenoy

    Department of Chemical Engineering, University of California, Santa Barbara, CA 93106, USA

    Materials Research Laboratory, University of California, Santa Barbara, CA 93106, USA

    ,
    Chelsea ER Edwards

    Department of Chemical Engineering, University of California, Santa Barbara, CA 93106, USA

    Materials Research Laboratory, University of California, Santa Barbara, CA 93106, USA

    ,
    Matthew E Helgeson

    Department of Chemical Engineering, University of California, Santa Barbara, CA 93106, USA

    Materials Research Laboratory, University of California, Santa Barbara, CA 93106, USA

    &
    Megan T Valentine

    *Author for correspondence:

    E-mail Address: valentine@engineering.ucsb.edu

    Materials Research Laboratory, University of California, Santa Barbara, CA 93106, USA

    Department of Mechanical Engineering, University of California, Santa Barbara, CA 93106, USA

    Published Online:https://doi.org/10.2144/btn-2021-0009

    Abstract

    3D printing holds potential as a faster, cheaper alternative compared with traditional photolithography for the fabrication of microfluidic devices by replica molding. However, the influence of printing resolution and quality on device design and performance has yet to receive detailed study. Here, we investigate the use of 3D-printed molds to create staggered herringbone mixers (SHMs) with feature sizes ranging from ∼100 to 500 μm. We provide guidelines for printer calibration to ensure accurate printing at these length scales and quantify the impacts of print variability on SHM performance. We show that SHMs produced by 3D printing generate well-mixed output streams across devices with variable heights and defects, demonstrating that 3D printing is suitable and advantageous for low-cost, high-throughput SHM manufacturing.

    Method Summary

    We investigate the use of 3D printing to create staggered herringbone mixers (SHMs) and show that such devices generate well-mixed output streams across devices with variable heights and defects. This demonstrates that 3D printing is suitable and advantageous for low-cost, low-effort, high-throughput micromixer manufacturing.

    Microfluidic mixers such as the staggered herringbone mixer (SHM) [1] promote eddy-like mixing of laminar flow streams, avoiding prohibitively long channel lengths and enabling applications in drug delivery and discovery [2], chemical synthesis [3], sample concentration [4] and biological analysis [5–7]. In SHMs, asymmetric herringbone grooves embedded in the floor or ceiling of the rectangular channel cause transverse flow within fluid streams, promoting mixing by increasing local vorticity (Figure 1) [1,8]. Alternating the grooves' offset between cycles increases flow irregularity, further mixing the two streams [9]. The SHM's mixing proficiency and mechanism has been characterized extensively [8,10] due to its efficiency and ease of design compared with other grooved micromixers [11]. Their low shear flow properties [6] and ability to circulate flow within the channel [4] make SHMs particularly advantageous for biomedical applications.

    Figure 1. Schematic of one cycle of the staggered herringbone mixer (SHM), with dimensions labeled in the inset.

    h/H ∼ 0.36, w/d ∼ 0.45, and a ∼ 95°. See Supplementary Materials for additional information.

    Despite these advantages, the widespread use of SHM devices for on-chip diagnostics [4,6,12] is hindered by the reliance on time-intensive photolithography-based fabrication to generate molds for polydimethylsiloxane (PDMS)-based microfluidic devices. Photolithography requires cleanroom training and costly equipment and reagents [13]. Consequently, 3D printing is emerging as an attractive substitute for photolithography due to its comparative affordability, simplicity [14], and the ease of fabricating multilevel designs [13]. 3D printing also enables rapid prototyping by reducing typical fabrication times from days to hours [14].

    To establish the use of 3D printing to generate molds for PDMS-based microfluidic devices such as the SHM micromixer, experimental validation of printer and device performance is required. For current 3D printers [15], the minimum feature size is significantly larger than that provided by photolithography, print-to-print variability is expected and random disfigurations can occur, all potentially influencing device performance. Moreover, there are often discrepancies between the targeted and actual dimensions of the 3D printed parts [16]. Unfortunately, not only is information concerning such limitations generally unavailable for any given 3D printer model, but the lack of reporting on chip-to-chip variability is a known barrier for large-scale manufacturing of microfluidic devices [17]. Thus, understanding how print quality influences pattern transfer and mixing performance is especially important for low-cost, low-effort manufacturing that maintains efficacy, as required for microfluidic diagnostic platforms.

    To experimentally investigate print quality, we printed a series of raised channels, with heights and widths ranging from ∼25 to 700 μm, using a Miicraft Ultra 50 3D digital light processing (DLP) printer, with a horizontal and vertical resolution of ∼30 μm and ∼5 μm, respectively, and using Resin Works 3D Master Mold Resin for PDMS. We sized the 3D-printed parts by imaging with a Keyence VHX-5000 microscope (Supplementary Figure 1), and found them to be consistently smaller than the programmed design dimensions, with greater discrepancies in height than in the planar dimensions. These differences were consistent across all printed molds, so the programmed design dimensions could be calibrated to achieve printed parts with the correct size (Supplementary Figure 2).

    This calibration critically informed mold design, which was performed in Solid Works. Features with design dimensions ≥120 μm (∼100 μm actual dimension) were reproducibly printed, whereas features with smaller design dimensions were often deformed or occasionally missing. We then printed a series of herringbone grooves with a fixed width w of 100 μm and herringbone spacing d that varied from 50 to 500 μm (Figure 1). Printed grooves with designed d < 300 μm (∼220 μm real spacing) frequently fused with the adjacent herringbones, with increasing severity for decreasing d (Supplementary Figure 1A & B). We note that these disfigurations existed within the molds themselves, not only the PDMS replicates. Thus, we established a minimum feature size of 100 μm and fixed d = 300 μm in the design of SHM molds for further study.

    These constraints required SHM features in 3D-printed molds to differ somewhat from the geometries recommended by prior work [18], which find mixing to be most effective near w/d ∼1 and w = 50 μm. By contrast, our mixer has w/d ∼ 0.45 and w = 100 μm. However, we achieved a ratio of herringbone height h to channel height H, h/H ∼ 0.36, and herringbone angle a ∼ 95°, the latter of which was informed by computational work that suggested that angles in the range of ∼90°–105° produce optimal transverse flow [1,19]. Ten herringbones per each half cycle were included in accordance with prior experimental studies [10]. To compensate for the larger feature sizes and spacing requirements of 3D-printed molds and prevent the microchannel length from being excessively long, we limited the number of mixing cycle elements to five on a single chip (total channel length ∼41 mm) (Supplementary Figure 3). We then investigated a range of flow rates to produce well-mixed streams with these larger dimensions.

    To demonstrate the mixing capabilities of PDMS-based micromixer devices generated using the 3D-printed SHM molds, we examined the mixing performance of two exemplar devices generated with two identically designed, but individually printed, molds. In detail, PDMS, obtained as commercially available Sylgard 184 (Dow Chemical, MI, USA), was mixed at a 10:1 ratio of PDMS to cross-linker, poured in the treated molds, degassed for approximately 2 h, then cured at 60–80°C for 4 h and removed from the mold. We did not observe qualitative deviations in the PDMS replicates formed from either mold, consistent with the well-established high-fidelity pattern transfer from 3D-printed molds to PDMS [20]. Each device was tested at flow rates from 0.5 to 20 μL min-1, corresponding to Peclet numbers Pe = 200–8000, a useful range for low shear-stress mixing of biological materials [21]. The Peclet number describes the ratio of advective to diffusive transport and is given by Pe = uL/D where u is the average velocity defined by u = q/WH and q is the volumetric flow rate, L is a characteristic length scale given here by W/2 and D is the Stokes–Einstein diffusivity.

    For each test, a 0.30 μM aqueous solution of 70 kDa fluorescein isothiocyanate (FITC)-dextran was delivered into Input 1, while neat water was delivered into Input 2 at an equivalent flow rate. The mixing of the two streams under steady-state flow conditions was observed via fluorescence microscopy (see Supplementary Materials for details). Representative intensity maps of the flow fields along a cross-section of the channel after one, three and five mixing cycles are shown in Figure 2. For quantitative analysis, the intensity was normalized by dividing by the maximum value (Supplementary Figure 4), and its coefficient of variation (CV), defined as the standard deviation divided by the mean, was calculated (Figure 3 & Supplementary Figure 5). We consider any profile with a CV < 0.1 to be well mixed [10].

    Figure 2. Heat maps showing fluorescence intensity across the SHM channel cross-section after one, three and five cycles of mixing for Pe = 200 (left) and Pe = 8000 (right).

    The color scale indicates the presence of fluorescein isothiocyanate (FITC)-dextran, introduced through Inlet 1 and mixed with neat water introduced through Inlet 2. The dark blue regions indicate low fluorescence signal (i.e., mostly water), whereas red regions indicate high FITC signal; uniform light blue/green regions indicate homogeneous mixing. At Pe = 200 streams appeared homogeneous and well-mixed after three cycles, while at Pe = 8000 striations remained even after five cycles.

    Devices generated using 3D-printed SHM molds demonstrate good mixing performance. The number of cycles required to achieve good mixing depends on Pe, as expected. At Pe = 200, where diffusion is prominent, only three mixing cycles are required to achieve CV < 0.1, and only four to five mixing cycles are needed at Pe values of 800–6000 (Figure 3). We could not achieve good mixing for Pe > 6000 with this design due to the limited channel length. We observed a modest decline in mixer performance compared with that of devices with smaller feature sizes achieved via photolithography, where it is possible to achieve CV < 0.1 with two mixing cycles at Pe = 625 and with five cycles at Pe = 6250 [10].

    Figure 3. Comparison of intensity CV values for each cycle of mixing at various Pe values for (A) Device 1 and (B) Device 2.

    In both cases well mixed streams with CV < 0.1 black dashed line) are achieved for Pe < 6000. Error bars were calculated via propagation analysis as described in the Mixing Analysis section in the Supplemental Methods.

    Finally, we examined the effects of print-to-print variation on mixing by comparing the results obtained using two devices containing chipped herringbones, with the second device containing twice as many defects as the first, and modest differences in herringbone properties (Supplementary Tables 1–4). Device 2 exhibits larger CV values across almost all Pe values after one and two cycles of mixing; however, these differences vanish after three to five cycles of mixing (Supplementary Figure 5). That print-to-print variability does not compromise performance suggests that rigorous measurements of each mold are not necessary beyond the initial characterization of 3D printer capabilities and 3D printing can deliver SHM devices that yield reproducible results over multiple prints, even in the presence of defects.

    In summary, we introduced methods to characterize and calibrate 3D printer outputs and adapt the SHM geometry to enable 3D printing of replica molds. We demonstrated good mixing performance despite the modified dimensions and in the presence of print-to-print variation and defects. This method provides significant advantage for applications that benefit from rapid, low-effort manufacturing.

    Supplementary data

    To view the supplementary data that accompany this paper please visit the journal website at: www.future-science.com/doi/suppl/10.2144/btn-2021-0009

    Acknowledgments

    The authors thank David Bothman and Daniel Magnuson for technical assistance with printing.

    Author contributions

    VJ Shenoy and CER Edwards designed 3D prints, fabricated devices and collected experimental data. MT Valentine, ME Helgeson and CER Edwards supervised the project. All authors designed experiments, analyzed and interpreted data, and wrote the manuscript.

    Financial & competing interests disclosure

    This work was supported by the MRSEC Program of the National Science Foundation under award no. DMR 1720256 (IRG-3). The authors acknowledge the use of the Microfluidics Laboratory within the California NanoSystems Institute, supported by the University of California, Santa Barbara and the University of California, Office of the President. VJ Shenoy acknowledges support of the MRSEC-sponsored RISE and FLAM programs. CER Edwards also acknowledges support from the Department of Defense through the National Defense Science and Engineering Graduate Fellowship program. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

    No writing assistance was utilized in the production of this manuscript.

    Open access

    This work is licensed under the Attribution-NonCommercial-NoDerivatives 4.0 Unported License. To view a copy of this license, visit http://creativecommons.org/licenses/by-nc-nd/4.0/

    References

    • 1. Stroock AD, Dertinger SKW, Ajdari A et al. Chaotic mixer for microchannels. Science 295(5555), 647–651 (2002).
    • 2. Jia X, Wang W, Han Q et al. Micromixer based preparation of functionalized liposomes and targeting drug delivery. ACS Med. Chem. Lett. 7(4), 429–434 (2016).
    • 3. Kim DJ, Oh HJ, Park TH et al. An easily integrative and efficient micromixer and its application to the spectroscopic detection of glucose-catalyst reactions. Analyst 130(3), 293–298 (2005).
    • 4. Lee H-Y, Voldman J. Optimizing micromixer design for enhancing dielectrophoretic microconcentrator performance. Anal. Chem. 79(5), 1833–1839 (2007).
    • 5. Lee TY, Hyun K-A, Kim S-I, Jung H-I. An integrated microfluidic chip for one-step isolation of circulating tumor cells. Sens. Actuators B Chem. 238, 1144–1150 (2017).
    • 6. Lehmann M, Wallbank AM, Dennis KA et al. On-chip recalcification of citrated whole blood using a microfluidic herringbone mixer. Biomicrofluidics 9(6), 064106 (2015).
    • 7. Stott SL, Hsu C-H, Tsukrov DI et al. Isolation of circulating tumor cells using a microvortex-generating herringbone-chip. Proc. Nat. Acad. Sci. USA 107(43), 18392 (2010).
    • 8. Kee SP, Gavriilidis A. Design and characterisation of the staggered herringbone mixer. Chem. Eng. J. 142(1), 109–121 (2008).
    • 9. Wiggins S, Ottino JM. Foundations of chaotic mixing. Philos. Trans. Royal Soc. 362(1818), 937–970 (2004).
    • 10. Williams MS, Longmuir KJ, Yager P. A practical guide to the staggered herringbone mixer. Lab on a Chip 8(7), 1121–1129 (2008).
    • 11. Nam-Trung N, Zhigang W. Micromixers-a review. J. Micromech. Microeng. 15(2), R1 (2005).
    • 12. Lund-Olesen T, Dufva M, Hansen MF. Capture of DNA in microfluidic channel using magnetic beads: Increasing capture efficiency with integrated microfluidic mixer. J. Magnetism Magnetic Mater. 311(1), 396–400 (2007).
    • 13. Hwang Y, Paydar OH, Candler RN. 3D printed molds for non-planar PDMS microfluidic channels. Sens. Actuators A. 226, 137–142 (2015).
    • 14. Gross BC, Erkal JL, Lockwood SY et al. Evaluation of 3D printing and its potential impact on biotechnology and the chemical sciences. Anal. Chem. 86(7), 3240–3253 (2014).
    • 15. Chen C, Mehl BT, Munshi AS et al. 3D-printed microfluidic devices: fabrication, advantages and limitations-a mini review. Anal. Meth. 8(31), 6005–6012 (2016).
    • 16. Alsoufi MS, Elsayed AE. Surface roughness quality and dimensional accuracy-a comprehensive analysis of 100% infill printed parts fabricated by a personal/desktop cost-effective FDM 3D printer. Mater. Sci. Appl. 9(1), 11–40 (2018).
    • 17. Volpatti LR, Yetisen AK. Commercialization of microfluidic devices. Trends Biotech. 32(7), 347–350 (2014).
    • 18. Chen X, Wang X. Optimized modular design and experiment for staggered herringbone chaotic micromixer. Int. J. Chem. Reactor Eng. 13(3), 305–309 (2015).
    • 19. Ansari MA, Kim K-Y. Shape optimization of a micromixer with staggered herringbone groove. Chem. Eng. Sci. 62(23), 6687–6695 (2007).
    • 20. Bonyár A, Sántha H, Varga M et al. Characterization of rapid PDMS casting technique utilizing molding forms fabricated by 3D rapid prototyping technology (RPT). Int. J. Mater. Forming 7(2), 189–196 (2014).
    • 21. Shen F, Li X, Li PC. Study of flow behaviors on single-cell manipulation and shear stress reduction in microfluidic chips using computational fluid dynamics simulations. Biomicrofluidics 8(1), 014109 (2014).